Appendix c: hi-4416 operating protocols, Communication protocol, Probe data format – ETS-Lindgren HI-4416 Numeric EMF Readout Unit User Manual

Page 29

Advertising
background image

www.ets-lindgren.com

HI-4416 Operating Protocols

29

Appendix C: HI-4416 Operating Protocols

Communication between the probe and the HI-4416 System Read Out is carried

out via an ASCII data string. The probe requires a dual fiber optic cable and only

responds when commanded. The HI-4416 continuously (at 7.6 Hz) sends the

probe a command for data and waits for a response. If a key that requires

information from the probe is pressed, the HI-4416 will send the appropriate

command to the probe and wait for a response.

C

OMMUNICATION

P

ROTOCOL

Data Type:

RS-232 Serial

Data Mode:

Asynchronous

Word Length:

7 bits

Parity: odd

Stop Bits:

1

Data Rate:

9600 baud

P

ROBE

D

ATA

F

ORMAT

The data sent to the HI-4416 is formatted as SDxx.xxuuugggobaaat:

S

Start Character (":")

D

Type Indicator ("D" = Controller, "#" = Listen Only)

xx.xx

Probe Reading (4 digits plus floating decimal)

uuu

Units of Measure (see next page for valid unit indicators)

ggg

Recorder Output/Bar Graph (0 - 255)

o

Over Range Indicator ("N" = Normal, "O" = Over Range)

b

Battery Status ("N" = Normal, "W" = Warning, "F" = Fail)

aaa

X-,Y-,Z-Axis (respectively) enable flag

("E" = Enabled, "D" = Disabled)

t

Terminating Character (Return)