Operation

www.ets-lindgren.com

Once the system readout is turned on, all segments of the LCD will activate for

two seconds, the version of software installed will appear, followed by the current

range of the probe connected to the readout. The HI-4416 then switches to

normal operation (measurement). The LCD readout displays the observed value

and the units of measure.

The bar graph along the top of the LCD window presents an analog

approximation of the measured field. Each bar graph segment represents five

percent of the full-scale reading in the current range. The bar graph is updated

7.6 times per second while the digital display is updated 1.9 times per second.

When the measured field strength is below 5% of full scale for the range in use,

the "Under Range" indicator at the left end of the bar graph appears and the

display will flash on and off. When possible, switch the range to permit a field

strength reading that does not trigger the "Under Range" indicator. If the field

strength exceeds full scale, the "Over Range" indicator at the right end of the bar
graph will appear and the characters “OL” will appear on the display. Select the

next appropriate range.