Operation

21

The maximum value is cleared when the HI-4416 is powered up, when the Max

key is released and when the unit of measure is changed. The Clear button does

not affect this reading.

If the maximum reading is over full scale for the range, 'OL' will appear

on the display along with the over range indicator.

Zero

Sends a zero command to the probe.

Selecting this command zeroes all ranges and axes. This mode establishes a

baseline for measurements by sampling all axes and ranges and subtracting

those values from each subsequent measurement.

Batt

Displays probe battery voltage and temperature. Refer to the manual for

your probe to determine if it is capable of communicating battery status.

Pressing this key with the probe connected and armed causes the LCD to display

a small battery symbol on the left side of the readout and the battery voltage of

the probe. After two seconds, the readout will also display probe temperature

in °F. After another two seconds, the readout returns to the measurement mode.

Compare the voltage reading you obtain with that stated in the probe manual.

When the probe's battery voltage decreases below a preset limit, the battery

symbol will blink; this indicates that the battery needs charging. If the battery

voltage is allowed to drop below that required for proper device operation, the

display will be blank.

This keypad can also be used to determine the battery status of the HI-4416.

When the battery symbol is flashing, toggle the probe's Arm/Off switch to Off, if

the symbol still blinks, then the battery in the system readout needs charging.

Units

Instructs the probe to change the units of measure, and displays the new

units.

Continue to press the Units keypad until the desired unit of measure appears on

the display. Just as for the Range command, the available units of measure

depend upon which probe is connected to the system readout; consult your

probe manual.