Installation

MODEL 94456 CURRENT PROBES

6

© EMC TEST SYSTEMS, L.P. – MARCH 2002

REV B – PN 399265

conditions, several measurements taken along the line will

provide a complete picture of the RF current distribution

and amplitude.

INSTALLATION

The window (aperture) of the probe will accommodate

cables up to a maximum outside diameter of 4.0 inches.

After placing the probe around the conductor(s) to be

measured, the probe jaws should be carefully locked. If this

is not done, inadequate shielding or incorrect air gap will

result and the measurement will not be accurate.

The connecting cable used between the current probe and

the EMI test equipment must have a 50 Ohm characteristic

impedance and matching cable connectors. The current

probe is calibrated for use only with a 50 Ohm load.

Therefore, the EMI test equipment must have a 50 Ohm

input impedance. Precautions regarding minimum bending

radius should be observed when installing and using the

cable. For long cables and at high frequencies, cable loss

may also be a factor. Care should be taken to use low loss

cables and to perform cable loss corrections if necessary.

The probe rejection of any external pickup from conductors

not passing through the window is better that 60 dB. The

presence of very strong magnetic fields will likely have an

effect on probe sensitivity. Care must be taken not to place

the unit close to permanent magnets or the magnetic field

structures of motors or generators.