MODEL 94456 CURRENT PROBES

8

© EMC TEST SYSTEMS, L.P. – MARCH 2002

REV B – PN 399265

PRECAUTIONARY MEASURES

When measuring uninsulated conductors use extreme care

when installing the current probe and taking measurements.

If possible, de-energize the test sample during assembly

and disassembly of the setup. Also, arrange to center the

test conductor in the current probe window for additional

voltage breakdown protection.

Do not permit the uninsulated current probe connector and

cable connectors to come in contact with the ground plane

or other nearby conductors. This will prevent possible

measurement error due to ground loops, and will avoid

danger from high voltages.

Ensure that the 50 Ohm load is capable of safely dissipating

the incurred power. Should the load become disconnected,

the developed voltage will be come much greater and may

be very dangerous.