MODEL 94456 CURRENT PROBES

Operation

© EMC TEST SYSTEMS, L.P. – MARCH 2002

11

REV B – PN 399265

interference in terms of dB above one microamp-per-

megahertz at meter input.*

Example:

Frequency is 100 kHz; Step (1) measurement is 41 dB

above one microvolt-per-megahertz. For example, suppose

the transfer impedance of the current probe was 8.0 dB

below one Ohm at 100 kHz. Then, as outlined in Step (2):

41 dB + 8.0 dB = 49 dB above one microamp-per-

megahertz at meter input.

2

This result is beyond the limit of 46.2 dB above one

microamp-per-megahertz .

In Terms of Microampere in Test Sample Lead

(CW Conducted Measurements)

1. Adjust the EMI test equipment for standard gain and

make a measurement of the CW signal (voltage output

from current probe) in terms of microvolts at meter

input. Use procedures outlined in the EMI test

equipment instruction manual.

2. Divide the microvolt measurement of Step (1) by the

transfer impedance in Ohms at the test frequency. The

result is the value of conducted CW signal in terms of

microamperes in the test sample lead.

Example:

Frequency is 3.0 kHz; Step (1) Measurement is 150

microvolts. For example, suppose the transfer impedance of

2

The term “at meter input” as used in the MIL-I-26600 and MIL-I-6181D specifications refers to the

current in the test sample lead.