Installation Manual
PEAK HOLD
OFF
NORMAL
BI-DIR
AUTO
Peak hold. Used to determine, display, and print the greatest net weight read during a
weighing cycle. The weighing cycle ends when a print command is executed (AUTO setting)
or when the peak weight is cleared by pressing ZERO or PRINT. Press GROSS/NET to
display gross weight data when using the peak hold function.
OFF
Peak hold function is off.
NORMAL Positive peak, manual reset. Greatest net weight is held in memory until the
weight is removed from the scale and either the ZERO or PRINT key is pressed.
BI-DIR
Bi-directional peak, manual reset. Same as NORMAL, but peak value can be
either positive or negative, determined by absolute value.
AUTO
Positive peak, auto print, auto reset. Automatic print occurs when scale load is 0
± 10 display divisions and at standstill. Following the print command, the peak
value is cleared and reset automatically.
CALIBR
WZERO
WVAL
WSPAN
WLIN
REZERO
See Level 4 submenu descriptions in Table
29.
SCALES Menu
Parameter
Choices
Description
Table 3-2. SCALES Menu Parameters (Continued)