1 operation – Rice Lake Ishida iPC-WP Portable Bench Scales User Manual
Page 18

15/32
IPC-WP Series Service Manual
2.8
F Mode – Setting Measuring Conditions and E2ROM Clear
Note
• All data has been fixed according to country specifications.
• Altering data may not conform to weighing and measuring tests for a country.
Reference
Reference
• See 1.6 “C1 Mode-Country No. Setting” for Measurement Conditions Setup
Table.
2.8.1 Operation
Operation
Display
1. Stating
Test
2. Setting measuring conditions
2.1 Press the TARE key for three times
2.2 Press the [ *] key → The least significant digit blinks.
Whenever the [ *] key is pressed, the required digit can be
selected.
2.3 Whenever the [ *] key is pressed, the figure of the digit which
has blinked does the increment.
3. E2ROM
clear
・
Any scene of F mode is possible.
Press the ZERO key and [ * ] key. → "EP-C" then "C1-1" is
displayed.
↓
Note
At this point, writing in E2ROM of change data in F
mode and the default value to E2ROM clear has
not been completed. Push the memory switch.
Note
E2ROM clear initializes Country No., Scale No.,
Zero, and Span adjustment values. It is necessary
to set it again.
4. Writing in E2ROM
・
Push the memory switch to memorize data.
Note
After E2ROM is cleared, if writing is performed in
E2ROM without setting Country No., Scale No. nor
Zero/Span adjustment value, and the power is
turned on, "Err1" is displayed and it is not possible
to use the machine. Set C1, C2, and C3 again.