Understanding the LCD messages
Welch Allyn MicroTymp3 portable tympanometric instrument
Open
The test has not begun since a valid ear cavity has not been detected.
Possible causes:
•
Instrument is not in an ear
•
Probe tip is not completely sealed in an ear
•
Perforated tympanic membrane
•
Ear with patent tympanostomy tubes
•
Extremely large canal (> 2.5 cc)
Possible solutions:
•
Reposition the probe tip
•
Perform otoscopy to check for perforated membrane or patent tympanostomy tubes
Test
The TEST message indicates that the test has started. Test data will begin to appear
immediately.