Maintenance and troubleshooting
Welch Allyn MicroTymp3 portable tympanometric instrument
1.
Test Pattern - The test pattern is used to confirm that the print head is functioning 
properly. If any of the print head elements are not functioning, a white line will appear 
vertically down the printout. A defect in the paper may also cause white lines or light 
printing. Repeat the test pattern to confirm any suspected printing problems. If paper 
advances but nothing prints out, check to be sure the paper is inserted properly
(see
2.
Software Version - The test pattern also includes the software version for the Printer/
Charger. 
3.
Normative Data Reference - The test pattern is followed by normative data for 
tympanometric characteristics for both the MicroTymp2 and MicroTymp3 Handle. This 
is provided along with the test pattern for the convenience of the user, not specifically 
as a functional check.