Interpreting 1000-hz results, Normative values, Pass – Welch Allyn MicroTymp 3 portable tympanometric instrument - User Manual User Manual
Page 38

38
Interpreting test results
Welch Allyn MicroTymp3 portable tympanometric instrument
Normative values
The table below presents normative values for 1000-Hz tympanometry for NICU babies
and full-term 2- to 4-week-old babies. [Margolis, R.H., Bass-Ringdahl, S., Hanks, W.D.,
Holte, K., Zapala, D.A. Tympanometry in Newborn Infants - 1 KHz Norms. J. Amer. Acad.
Audiol., 14, 383-392, 2003.]
Interpreting 1000-Hz results
Pass
A normal tympanogram will yield a PASS result. The displayed numerical value represents
the compensated admittance.
Condition which causes a PASS message:
•
Peak Ya greater than or equal to 0.6 mmho
Normative tympanometric values (1000 Hz)
Tympanometric Measurement
NICU
90% Range
Full-term
90% Range
Peak Ya
(Negative Tail Compensation)
0.6 - 2.7 mmho
0.6 - 4.3 mmho
Tympanometric
Peak Pressure (TPP)
-93 - +53 daPa
-133 - +113 daPa