Lamp and Battery Replacement

COVER REMOVAL AND REPLACEMENT

If the lamp or battery is to be replaced, it is necessary to first
remove the instrument cover:

1. Disconnect the charging transformer from AudioScope 3 and

turn the instrument “OFF” before beginning.

2. Insert a coin into the slot of the lock mechanism at the end of

the handle.

3. Turn the coin in either direction until the line under the word “OPEN”

on the lock is aligned with the recessed line on the housing.

4. Slide the cover back (away from the speculum end), until the

arrows on the instrument case align.

5. Lift the cover.

NOTE: Once the cover is removed, care should be taken not to
touch any of the internal components except as described in
this manual.

Step 2

Step 3

Step 4

Step 5

WANM-3999-COL AudioScope3 3/11/97 1:24 PM Page 14