Welch Allyn Audioscope 3 Portable Screening Audiometer - User Manual User Manual

Page 28

Advertising
background image

26

Operating Temperature Range: 59° to 95°F (15° to 35°C).

All specifications apply over this temperature range.

Frequencies: 500, 1000, 2000 and 4000 Hz ±3%.

Distortion: Less than 3% total harmonic distortion (as measured at

transducer input).

Rise/Fall Times: 60 ±40 msec. as measured between the -1 and

-20 dB points.

Tone Duration: 1.5 ±0.2 sec.

Pause Between Tones: Varied: between 1.0 and 2.0 seconds.

Sound Level: Sound levels were established by threshold loudness
balance method to TDH-39 receiver with MX41/AR cushion per
ANSI S3.6 -1969 Ref. Thresholds.

These levels were established in an independent study and are
equivalent to standard audiometric headphones.

5

±3 dB @ 500, 1000 and 2000 Hz ±4 dB @ 4000 Hz. This corre-
sponds to the following absolute sound pressure levels when the
instrument is coupled to an occluded ear simulator per ANSI Std.
No. S3.25-1979.

Actual Sound Level (dB re 20 uPa)

Freq. (Hz)

For 20 dB HL

For 25 dB HL

For 40 dB HL

500

29.9

34.9

49.9

1000

27.3

32.3

47.3

2000

32.7

37.7

52.7

4000

33.9

38.9

53.9

5. Bienvenue, G., Michael, P., Chaffinch, J. — “Reference Threshold Sound Pressure Levels for

the Welch Allyn Audioscope”, Acoustical Society of America, 1984