Welch Allyn Audioscope 3 Portable Screening Audiometer - User Manual User Manual

Page 17

Advertising
background image

Lamp and Battery

Replacement

15

LAMP REPLACEMENT

To remove old lamp:

1. Turn AudioScope 3 “OFF.”

2. LAMP MAY BE HOT—ALLOW TO COOL BEFORE

PROCEEDING.

3. Follow instructions for cover removal (page 14).

4. Depress the blue lamp ejector lever (see Figure 1 below). The

lamp will pop up.

5. Grasp lamp and remove.

Figure 1

To insert new lamp:

1. For best performance of Halogen lamps, remove any dirt or

fingerprints from glass with a clean cloth moistened with 70%
Isopropyl Alcohol before inserting lamp. Failure to follow clean-
ing procedures will result in permanent stains on the glass
with reduction in light output. Average lamp life is 20 hours of
on-time.

2. Rest lamp on lamp holder bracket with beveled end of lamp

toward fiber optic bundle (see Figure 1 above).

3. Push lamp into lamp holder.

4. Follow instructions for cover replacement (page 16).

NOTE: Replace with Welch Allyn No. 06200 lamp only. Other
lamps may cause damage to the instrument or may cause
inaccurate readings.

Lamp ejector lever

Fiber Optic Bundle
Beveled end of lamp

WANM-3999-COL AudioScope3 3/11/97 1:24 PM Page 15