27

Maximum Permissible Ambient Noise for 20 dB HL Screen

55 dB A or
48 dB — 500 Hz octave filter
48 dB — 1000 Hz octave filter
48 dB — 2000 Hz octave filter
48 dB — 4000 Hz octave filter

Maximum Permissible Ambient Noise for 25 dB HL Screen and
40 dB HL Screen

60 dB A or
53 dB — 500 Hz octave filter
53 dB — 1000 Hz octave filter
53 dB — 2000 Hz octave filter
53 dB — 4000 Hz octave filter

Lamp: 3.5v Halogen, 20 hr. average life

Battery: 3.5v nickel cadmium rechargeable — approximately
50 minutes of continuous operation between recharges.

CHARGING STAND SPECIFICATIONS

Dimensions:

Circuit
Diagram:

Model: 71123

Input: Models 71032, 71034, 71036 and 71040 Charging
Transformers

Output: Model 23300 AudioScope 3™ charging rate — 75 mA

Leakage Current: See instructions for 71032, 71034, 71036 and
71040 Charging Transformers.

Weight: 1 lb. 3 oz. (539 g).

4.5"

(11.4 cm)

2.75"

(6.99 cm)

4.25"

(10.8 cm)

330

75

2W

Pin Term

Sleeve Term

Power Jack

Charging Well

T

echnical Specifications

WANM-3999-COL AudioScope3 3/11/97 1:25 PM Page 27