Monitoring SpO
2
Warnings and Cautions — SpO
2
WARNING Always follow the manufacturer’s instructions for care and use of the
SpO
2
sensor.
WARNING The accuracy of the SpO
2
measurement can be affected by any of
the following:
•
the presence of significant amounts of dysfunctional hemoglobin, such as
carboxyhemoglobin or methemoglobin
•
the presence of concentrations of some intravascular dyes, sufficient to
change the patient’s usual arterial pigmentation
•
patient movement
•
patient conditions such as shivering and smoke inhalation
•
painted nails
•
poor oxygen perfusion
•
anemia or low concentrations of hemoglobin
•
hypotension or hypertension
•
severe vasoconstriction
•
shock or cardiac arrest
•
venous pulsations or sudden and significant changes in pulse rate
•
proximity to an MRI environment
•
moisture in the sensor
•
excessive ambient light, especially fluorescent
•
wrong sensor or sensor too tight
WARNING If there is any question of the accuracy of an SpO
2
measurement,
verify the measurement using another clinically accepted measurement method.