MAP Measurement Enable and Disable
Default setting: MAP ENABLED for neonate; MAP DISABLED for adult and pediatric.
1.
Depress
for 3 seconds. SET HOUR XX appears in the message window.
2.
Press
repeatedly until MAP ENABLED or MAP DISABLED appears in the
display window.
3.
Press
or
to enable or disable MAP measurement.
4.
To select the displayed state and return to normal operation, either do nothing for 10
seconds or press any button other than
or
. If you press a function button (such
as
), the monitor returns to normal operation with that function (
) activated.
For information about MAP measurements, see
Note
If you change the MAP enabled/disabled setting, refer to
Patient Type Affects MAP Defaults”