Nurse call – Welch Allyn 300 Series Vital Signs Monitor - User Manual User Manual
Page 61

Directions for Use
Chapter 4 Alarms and Alerts
57
Nurse Call
The monitor can be connected to a Nurse Call system through a customized cable that
connects to the Nurse Call connector. When the cable is connected and operational, the
monitor immediately notifies the Nurse Call system when a patient alarm occurs.
To connect the monitor to a Nurse Call system, you must have a cable (Welch Allyn part
number 008-0634-XX or equivalent) that has been adapted to your Nurse Call system. This
cable can be built or customized by a biomedical engineering department or other
appropriate technical resource, using the specification described on
WARNING Although the Nurse Call option enables remote notification of an
alarm condition, it is not intended to replace appropriate bedside patient
surveillance by trained clinicians.
Note
When a patient alarm occurs, pressing
silences the monitor alarm tone and
the Nurse Call alarm for 90 seconds, but does not suspend the visual alarm
indicator on the monitor.
Nurse Call Cable
Connector