Default Settings
Default Selections
Patient Mode Default
__ Adult
__ Pediatric
__Neonatal
Alarm Tone Level Default
__ Low
__ Medium
__ High
ECG Bandwidth Default
__ Monitor
__ Extended
ECG Bandwidth refers to the frequency range used to display and print patient data.
•
Monitor bandwidth is less susceptible to artifact such as patient motion.
•
Extended bandwidth, although more susceptible to artifact, can provide for more
detailed data analysis.
HR/PR Tone Level Default
__ Off
__ Low
__ Medium
__ High
Monitor-Mode Bandwidth
Adult
0.5 Hz to 40 Hz
Pediatric/Neonate
0.5 Hz to 80 Hz
Extended-Mode Bandwidth
Adult
0.05 Hz to 40 Hz
Pediatric/Neonate
0.05 Hz to 80 Hz