126

Welch Allyn Propaq LT Vital Signs Monitor

Default Settings

Default Selections

Patient Mode Default

__ Adult
__ Pediatric
__Neonatal

Alarm Tone Level Default

__ Low
__ Medium
__ High

ECG Bandwidth Default

__ Monitor
__ Extended

ECG Bandwidth refers to the frequency range used to display and print patient data.

Monitor bandwidth is less susceptible to artifact such as patient motion.

Extended bandwidth, although more susceptible to artifact, can provide for more
detailed data analysis.

HR/PR Tone Level Default

__ Off
__ Low
__ Medium
__ High

Monitor-Mode Bandwidth

Adult

0.5 Hz to 40 Hz

Pediatric/Neonate

0.5 Hz to 80 Hz

Extended-Mode Bandwidth

Adult

0.05 Hz to 40 Hz

Pediatric/Neonate

0.05 Hz to 80 Hz