Cradle, Battery – Welch Allyn 901061 Propaq LT Vital Signs Monitor - User Manual User Manual

Page 161

Advertising
background image

Directions for Use

Specifications

157

Battery

Cradle

Characteristic

Specification

Type

7.4 V, rechargeable, lithium ion. Field

-

replaceable by service technician.

Capacity

> 1800 mA

-

hr minimum

Charger

Contained in the cradle. Batteries are charged in place in the monitor.

Fuse rating

5A, 125V – User cannot access the fuse.
Note: Internal electronic overload circuitry is used as the primary method of protection. This
circuit resets itself when an overload is removed.

Recharge time

3 hours at 25° C (typical), with monitor either on or off.

Storage

< 30 days

-

4°F to 122°F

-

20°C to 50°C

30 days to 90 days

-

4°F to 104°F

-

20°C to 40°C

> 90 days

-

4°F to 86°F

-

20°C to 30°C

NOTE: For best charge retention during extended storage, remove the battery from the monitor.

Characteristic

Specification

Functional

Capacity

One monitor with internal battery.

Mounting

Tabletop or bed rail use. Can be attached to walls, IV poles, etc. with third

-

party

mounting brackets.

Protection Classifications

Duty cycle

Continuous

Type of protection against electric
shock

Type CF isolation exists between the monitor installed in the cradle and the
cradle DC power input, regardless of type of DC power input.
The line

-

operated power adapter accessory is a type CF medical

-

grade power

supply.
Type CF isolation exists between the monitor installed in the cradle and the
cradle USB connector.

Degree of protection against harmful
ingress of water

IPX1

Method of cleaning

Not suitable for autoclaving. See

“Inspecting and cleaning the monitor and

accessories”

on page 140.

Flammable anesthetics

Not suitable for use with flammable anesthetics.

Environmental

Operating temperature

0° to 40° C

Shipping and storage temperature

-

20° to 60° C

Operating altitude

-

500 to 15,000 feet (

-

152 to 4,572 m)

Shipping and storage altitude

-

2,000 to 40,000 feet (

-

610 to 12,192 m)

Operating relative humidity

5% to 95%, noncondensing

Shipping, storage relative humidity

5% to 95%, noncondensing

Shock

50 g