133

ParamSet Settings

ParamSet Enable

(Yes)
(No)

If ParamSet is enabled, the clinician can quickly and easily change the current alarm limits
for any vital sign. If you choose to enable ParamSet, you can then define the percentage
by which the alarm limit for each vital sign is changed.

To adjust alarm limits quickly with ParamSet

When the ParamSet message appears, press

twice.

Example

The ParamSet value for the Upper HR/PR alarm limit is configured to be 15%.

Patient Jane’s Upper HR/PR alarm limit is set to 100. At the default alarm settings, Jane
triggers an Upper HR/PR alarm with a heart rate of 103. Her clinician, knowing that a heart
rate of 103 is acceptable for Jane, wants to quickly set a higher Upper HR/PR alarm limit
for this patient.

The monitor displays the message, “Press Select to ParamSet the upper HR/PR limit to
118.” The clinician presses

to get to the HR/PR control screen and immediately

presses

again to accept the ParamSet adjustment. Jane’s alarm limit for Upper HR/PR

is immediately changed to 118, and the main vital-signs screen is again displayed.

(103 + 15% of 103 = approximately 118)

Note

If you select

(No), disregard the rest of the ParamSet settings.

HR/PR ParamSet %

Upper Limit +

20% (5, 10, 15, 20, 25)

Lower Limit -

20% (5, 10, 15, 20, 25)

Resp ParamSet %

Upper Limit +

5% (5, 10, 15, 20, 25)

Lower Limit -

5% (5, 10, 15, 20, 25)

SpO

2

ParamSet %

Upper Limit +

5% (5, 10)

Lower Limit -

5% (5, 10)

NIBP ParamSet %

Systolic

Upper Limit +

15% (5, 10, 15, 20, 25)

Lower Limit -

15% (5, 10, 15, 20, 25)

Diastolic

Upper Limit +

15% (5, 10, 15, 20, 25)

Lower Limit -

15% (5, 10, 15, 20, 25)

Mean

Upper Limit +

10% (5, 10, 15)

Lower Limit -

10% (5, 10, 15)