Wacs hl7 interface option, Outbound messaging – Welch Allyn Connectivity Server - Programmers guide User Manual
Page 62

58
Contacts and specifications
Welch Allyn Connectivity Server
WACS HL7 Interface option
Outbound messaging
Table 27. Configuration variables under HL7 tab
HL7 Manager page
Default setting
Alternate settings
Numeric data filter
Median (can be mean for
even number of samples)
Closest (most recent)
Numeric data filter for NIBP
Closest (most recent)
Median (can be mean for
even number of samples)
HL7 - General Settings page
HL7 versions supported
2.4
2.31, 2.3
Allow observations for multiple patients in one HL7 message
Off
On
Number of seconds to be retrieved when a new patient is
connected
7200
Field entry allowed
Allow multiple observations for a patient in one HL7 message
Off
On
Number of times the server will try to send an HL7 message to
client before disconnecting
5
Field entry allowed
Message delay (seconds) (for push)
420
Field entry allowed
HL7 - Push on Event Settings page
Each vital sign listed in page
Off
On
HL7 - Observation Identifiers page
Vital-sign labels
As listed in page
Field entry allowed
Table 28. Non-supported features
Simultaneous clients: WACS sends data to only one CIS
HL7 batch processing: WACS only sends data at a regular interval or in response to a query
Wildcards: WACS does not accept wildcards