Xml interface – Welch Allyn Connectivity Server - Programmers guide User Manual
Page 59

7
55
XML interface
The optional Third-Party Data Stream Interface option provides a means of moving patient
identification and patient numeric data between the Acuity Central Station and a facility’s
information system.
To receive XML messages from WACS, a CIS/HIS opens a TCP/IP socket and listens.
When the Welch Allyn Connectivity Server detects an event, it connects to the socket and
sends a message.
The interface provides the following parameters:
Table 24. XML interface parameters
Parameter
Content
Note
Command
Set
Reset
When the alarm starts.
When the alarm ends.
SourceID
[Patient ID or
Amended ID]
Acuity uses this as the unique patient identifier across
the Acuity network.
Sensitivity
Lethal
High
Parameter
ApplicationID
Welch Allyn
Text
[free-format text]
Alarm type. A valid alarm string (e.g., Heart Rate 250).
MessageID
Event ID
For WACS internal use only.
Timestamp
[time of the alarm]
GuestName
[Full name of the patient]
GuestPhysicalLocation
[Assigned location of the patient]
Unit and room number.
Parameter
[Model name and serial number of
the alarming monitor]
DestinationList
[Target recipient ID]
Present only if:
- Clinician Notifier option is licensed
- Event Delivery Mode = Escalation
- At least one Primary Respondent assigned