Alerts table – Welch Allyn EarlySense System 1.0 - Equivalent to model EverOn 1.0 - User Manual User Manual
Page 41

35
Alerts Table
The system uses the following alerts:
Message
Color
Description
High/Low Heart Rate:___
Red
Heart Rate exceeds thresholds.
High/Low Resp. Rate:___
Red
Respiration Rate exceeds thresholds.
Bed Exit
Sensitivity Settings:
1 (Low) – 6 (High)
Red
Bed Exit
Patient Out of Bed
For ____ Min
Red
System detected that patient is out of bed for a time
interval exceeding a preset threshold.
Patient Signal Lost
Red
System cannot detect patient’s signal.
Check patient and sensor location.
Extremely High Motion
Yellow
Patient Movement Level has been extremely high
during the last minute.
Patient Turn is OFF
Low Patient Movement
Yellow
Patient Movement Level has been very low during
nighttime (00:00-6:00). Consider setting Patient Turn
reminder ON.
Patient Turn is OFF
Air Mattress Motion
Detected
Yellow
Air Mattress was detected on the bed. Consider setting
Patient Turn reminder ON.