Audible and visual alert notification, Blue alerts – Welch Allyn EarlySense System 1.0 - Equivalent to model EverOn 1.0 - User Manual User Manual
Page 45

39
Thresholds can be set in the following ranges:
Lowest settable limit
threshold
Highest settable limit
threshold
Heart Rate
35 BPM
150 BPM
Respiratory Rate
8 Br/min
44 Br/min
Audible and Visual Alert Notification
When an alert is triggered an audible alert sounds and the corresponding alert parameter
on the monitor main screen will be shown with its background color corresponding to the
alert color, either red or yellow (Blue alerts are not connected to a specific monitoring
parameter). A message will appear on the bottom of the window detailing the nature of
the Alert (Figure 22) and the time passed since alert was raised
Figure 22: Alert Screen
Blue Alerts
Blue alerts are managed differently than yellow or red alerts. Blue alerts cannot be
suspended; their volume is set at maximum and is non-adjustable. Blue alerts appear only