15 infarction, Inferior infarction – Welch Allyn Means ECG Physicians Manual for CP Series Electrocardiographs - User Manual User Manual

Page 19

Advertising
background image

MEANS Physicians Manual


Welch Allyn

19

2.15 Infarction

This section of the program classifies an infarction according to location and estimates the
probability of its presence. The diagnosis of infarction is largely based on the presence and
duration of Q waves, Q/R ratios, and QS patterns. T and ST abnormalities are used in
statements on the age of the infarct (see section Repolarization).

The location of the infarction is determined by the leads in which the abnormalities are found.
The program distinguishes six locations: septal (lead V1, V2), anterior (V3, V4), lateral (V5,
V6), high lateral (I, aVL), inferior (aVF, III), and posterior (V1, V2). Lead II may be involved in
inferior infarction as well as in lateral infarction. Combined and more extensive infarcts will
generate infarct statements for more than one location (see section Combination of
statements).

Four degrees of probability are distinguished: “definite”, “probable”, “possible”, or “consider”.

Criteria on which a diagnosis of infarction has been made can also lead to other diagnoses,
like LBBB, RBBB, LVH and RVH. The choice between these possible diagnoses is based on
exclusion logic in the program, and in some situations, probabilities are adapted or criteria are
tightened.

Inferior infarction

An abnormal Q wave must be found in aVF and either II or III to even consider the diagnosis
of inferior infarction. As a rule the Q wave in aVF is shallower and shorter than that in III. Its
threshold to qualify as an infarct Q can therefore be lower.

Inferior infarction may produce a left (i.e., superior) axis deviation if one only considers the
ratio of upward and downward forces. In inferior myocardial infarction, however, this ratio is
shifted due to initial negativity (Q in aVF) whereas in ordinary left axis deviation it is
associated with deepening of the S wave in aVF.

Skip tests
if:

Q amplitude + R amplitude in aVF < 200 µV

or

Q amplitude in aVF

100 µV

or

Q amplitude in II

70 µV

and

Q amplitude in aVF

70 µV

or

Q amplitude in III

100 µV


Say:

“inferior infarct”

if:

Q duration

40 ms and Q/R ratio

0.3 in aVF


Say:

“probable inferior infarct”

if:

Q duration

40 ms and 0.2

Q/R ratio < 0.3 in aVF

or

30

Q duration < 40 ms and Q/R ratio

0.3 in aVF

or

Q duration in aVF

20 ms

and

Q duration

50 ms and Q amplitude > 300 µV in III


Say:

“possible inferior infarct”

if:

30

Q duration < 40 ms and 0.2

Q/R ratio < 0.3 in aVF

or

20

Q duration < 30 ms and Q/R ratio

0.3 in aVF

and

Q duration in III

40 ms

or

Q duration in aVF

20 ms

and

40

Q duration < 50 ms and Q amplitude > 300 µV in III

Advertising