Table 3. Accuracy of diagnostic interpretative statements on the CSE contour diagnostic
database vs. combined referee.
Diagnostic
Category
No. of
ECG
s
tested
Sensitivity;
%
Specificity:
%
Positive predictive
value %
Comments
Normal
381
98.4
51.9
98.2
LVH
181
82.0
89.9
94.8
RVH
52
40.1
100.0
100.0
BVH
51
69.6
98.2
97.0
AMI
170
89.1
89.0
98.0
IMI
273
78.7
89.6
96.5
MIX
73
79.6
87.9
95.7
VH+MI
31
81.1
96.0
96.8
HYPER
N/A
N/A
N/A
N/A
MI
N/A
82.9
91.5
97.9
References
1. Willems JL, Arnaud P, Van Bemmel JH, Degani R, Macfarlane PW, Zywietz C. Common
standards for quantitative electrocardiography: goals and main results. Methods Inf Med
1990;29:263-71.
2. Willems JL, Arnaud P, Van Bemmel JH, et al. A reference data base for multilead
electrocardiographic computer measurement programs. J Am Coll Cardiol 1987;10:1313-
21.
3. Willems JL, Abreu-Lima C, Arnaud P, et al. The diagnostic performance of computer
programs for the interpretation of electrocardiograms. N Engl J Med 1991;325:1767-73.
4. Willems JL. Common Standards for Quantitative Electrocardiography, 10
th
Progress
Report. Leuven: ACCO Publ, 1990.