Table 3. Accuracy of diagnostic interpretative statements on the CSE contour diagnostic
database vs. combined referee.

Diagnostic

Category

No. of

ECG

s

tested

Sensitivity;

%

Specificity:

%

Positive predictive

value %

Comments

Normal

381

98.4

51.9

98.2

LVH

181

82.0

89.9

94.8

RVH

52

40.1

100.0

100.0

BVH

51

69.6

98.2

97.0

AMI

170

89.1

89.0

98.0

IMI

273

78.7

89.6

96.5

MIX

73

79.6

87.9

95.7

VH+MI

31

81.1

96.0

96.8

HYPER

N/A

N/A

N/A

N/A

MI

N/A

82.9

91.5

97.9

References

1. Willems JL, Arnaud P, Van Bemmel JH, Degani R, Macfarlane PW, Zywietz C. Common

standards for quantitative electrocardiography: goals and main results. Methods Inf Med
1990;29:263-71.

2. Willems JL, Arnaud P, Van Bemmel JH, et al. A reference data base for multilead

electrocardiographic computer measurement programs. J Am Coll Cardiol 1987;10:1313-
21.

3. Willems JL, Abreu-Lima C, Arnaud P, et al. The diagnostic performance of computer

programs for the interpretation of electrocardiograms. N Engl J Med 1991;325:1767-73.

4. Willems JL. Common Standards for Quantitative Electrocardiography, 10

th

Progress

Report. Leuven: ACCO Publ, 1990.

Advertising