Temperature specifications
Characteristic
Specification
Range
0° to +50°C; 32° to +122°F
Displays
T1, T2, and
Δ
T
Probes
Compatible with YSI Series 400 and 700 probes. HP side panel
only compatible with YSI 400 and has HP connector.
Units
°C and °F selectable
Channel Accuracy
Temperature RangeTolerance
0° to +10°C±0.2°C
>10° to +50°C±0.1°C
+32° to +50°F±0.4°F
>50° to +122°F±0.2°F
Resolution
0.1°C or °F
Temperature Alarm Limit
Range (T1, T2)
32.0° to 122.0°F
0° to +50.0°C
Electrosurgery interference
suppression
Included in all models.