Set up the co2 display and alarm limits, Set up the co, Display and alarm limits – Welch Allyn 246 Propaq CS Vital Signs Monitor v.3.7X - User Manual User Manual

Page 50

Advertising
background image

46

Monitoring

Welch Allyn Propaq CS Vital Signs Monitor

Sidestream CO

2

accuracy decreases if additional tubing is connected to the sample

line. Avoid connecting additional tubing to the standard sample line.

Set up the CO

2

display and alarm limits

CO

2

monitoring is typically displayed as shown:

1.

To adjust the display, from the Main Menu press SpO

2

/CO

2

, CO

2

to access the first

CO

2

menu:

WARNING The exhaust port for sidestream CO

2

is an output for the expired

gases from the patient and any connected breathing apparatus. The exhaust port
is intended only for connection to gas collection equipment such as gas
scavenger devices (the device should comply with ISO 8835-3:1997 E). Do not
allow any other connection to the exhaust port.

WARNING If using a watertrap, the watertrap is disposable and should only be
used for a single patient. Do not reuse the watertrap for another patient.

Note

After you connect a mainstream CO

2

sensor or sidestream CO

2

watertrap, the

Propaq CS monitor displays the waveform briefly without a scale. It displays
WARM UP (for mainstream) or START UP (for sidestream) in the CO

2

numerics

window. After about 30 seconds, the monitor displays the CO

2

measurement and

waveform range.

NET

OFF

HR

MON

17:05:10

PRINT FAULT

II

BR

P

1

( 96)

BPM

mmHg

(M)

TEMP

100.4

98.6

1.8

97

121

79

S
D

P

2

( 15)

mmHg

(M)

25

9

S
D

NIBP

( 85)

mmHg

(M)

%

F

122

58

MCO

2

S

p

O

2

mmHg

38

Br/m

12

S
D

NIBP

ECG/

RESP

INVASIVE

PRESSURE

SpO

2

/

CO

2

SETUP

ADULT
ON NETWORK

T1
T2
ΔT

mV

cm

1

60

30

0

M

C

O

2

80

ALLEN KIM

MCO

2

mmHg

35

10

MCO

2

indicates mainstream CO

2

is active.

SCO

2

is displayed when sidestream CO

2

is active.

CO

2

is displayed if a mainstream sensor or sidestream watertrap is installed but not active.

ETCO

2

INCO

2

appears here when

displayed. The INCO

2

numeric value is only
displayed if it is in alarm or
if it is

≥ 7.5 mmHg (or 1kPa

or 1%).

Breath Rate is determined
from the CO

2

sensor.

Measured CO

2

levels are normally displayed as both a waveform and an ETCO

2

numeric value.

RANGE

mm/s

PREVIOUS

MENU

MORE