Power specifications – Welch Allyn 246 Propaq CS Vital Signs Monitor v.3.7X - User Manual User Manual

Page 116

Advertising
background image

112

Reference

Welch Allyn Propaq CS Vital Signs Monitor

Power Specifications

Characteristic

Specification

Mode of Operation

Continuous

Battery Pack Type

Sealed, gel-type lead acid

Battery Pack Capacity

Monitor only: 8 V, 2.7 Ampere-Hours; Monitor with Expansion Modules: 8 V, 5.4
Ampere-Hours

Battery Recharger Circuitry

Internal, powered by external power adapter

DC Input Power Required

12 to 28 V, 25 Watts

Input Fuse Rating

3A/250V, Type 2AG (0.57x 0.177 in)

Battery Recharge Time with
instrument on

Range of 8 hours to 12 hours typical, depending upon product configuration

Battery Recharge Time with
instrument off

Range of 6 hours to 8 hours depending upon product configuration

Recharge time until monitor is
usable, starting with discharged but
non-faulty battery

≤2 minutes typically (longer time required before NIBP, printer, and CO

2

are

available)

Low Battery Voltage and Operation

< 7.8 V: Caution message LOW BATTERY.
< 7.6 V: Caution messages LOW BATTERY, PRINTER DISABLED and LOW
BATTERY, NIBP DISABLED.
< 7.4 V: Equipment alert VERY LOW BATTERY
< 7.3 V: Equipment alert LOW BATTERY, HEATER DISABLED (MCO

2

)

< 7.0 V: Monitor automatically turns off.

Typical operating times on battery

Monitor only

2 hours

Monitor and SpO

2

(Baqpaq)

5 hours

Monitor with Expansion Module with
printer, SpO

2

and CO

2

Options

3 hours