Problem

Possible Cause

Solution

The water splash is
making a mess.

Excess water pressure
from the water line.

Ensure that the chamber
aerator is attached to the
bottom of the chamber.

The water pressure
from the eartip
seems too high or
too low.

The return water exit port
is clogged on the bottom
of the chamber.

Remove the chamber from the
faucet and dislodge any
foreign materials from the exit
port. Proceed with the normal
cleaning procedure.

There is an internal
malfunction.

With an eartip attached,
spray water from the handle
into a graduated cylinder or
measuring cup for exactly 20
seconds. If there was less than
90 mL or more than 110 mL of
measured water, contact your
Schering-Plough Animal Health
Office (see page 97).

Installation of eartip has
caused a problem.

Apply a new eartip with the
tab side pointing upwards.

There is leakage from the
chamber aerator and/or
adapter.

Tighten the aerator and/or
adapter with a pair of pliers.