Welch Allyn Propaq CS Monitor - User Manual User Manual

Page 16

Advertising
background image

8

English Update: Propaq CS Monitor

Welch Allyn Propaq CS Monitor

Summary of Alarm and Alert Keys and Acuity Central Station Messages

The following table summarizes Silence and Suspend behaviors and Acuity Central
Station messages for different combinations of current and previous Propaq monitor and
Acuity software versions.

Note

When your facility uses both older and newer Propaq monitors and Acuity
software versions, be aware that messages vary at the Acuity Central Station.
Some Acuity Central Station messages do not identify specific alarm/alert tone
silence or suspension states.

However, visual alarm indications still continue as usual at the Acuity Central
Station and at the Propaq monitors.

Propaq

Monitor

Version

Propaq Monitor

Key

Result of Propaq Monitor

Key Press

At Acuity Central Station, Result of Propaq

Monitor Key Press

Acuity 6.30 and prior Acuity 6.31 and later

Propaq CS 3.6X SILENCE or

Silences active alarm and alert
tones for 90 seconds at the
Propaq monitor and Acuity
Central Station.

All new alarms and alerts
interrupt the silence.

Alarms Suspended
Message

Only life-threatening
arrhythmia and apnea
alarms interrupt the
silence.

Alarms Suspended
Message

Only life-threatening
arrhythmia and apnea
alarms interrupt the
silence.

4 MIN SUSPND

Inhibits alarm and alert tones
for 4 minutes at the Propaq
monitor and Acuity Central
Station.

a

a.

If connected to the Acuity Central Station, life-threatening arrhythmias and apnea alarms interrupt the suspension.

Alarms Suspended
Message

Nurse Suspend 4 min
Message

ALL ALARMS

Indefinitely inhibits alarm and
alert tones at the Propaq
monitor and Acuity Central
Station, until someone resumes
tone capability.

a

Alarms Suspended
Message

Nurse Suspend Always
Message

Propaq CS 3.5X SUSPEND or

Suspends (for 90 seconds) the
capability for patient alarm and
equipment alert tones at the
Propaq monitor and Acuity
Central Station.

a

Alarms Suspended
Message

Alarms Suspended
Message

ALL ALARMS

Turns off all patient vital sign
alarm limits at the Propaq
monitor and Acuity Central
Station.

No visual or auditory alarm
indications occur until someone
resumes the limits.

a

Some Alarms Off,
Set Alarm Limits
Message

Some Alarms Off,
Set Alarm Limits
Message

Advertising