Emc compliance – Welch Allyn Propaq CS Monitor - User Manual User Manual

Page 19

Advertising
background image

Update to Directions for Use

English Update: Propaq CS Monitor

11

EMC Compliance

Special precautions concerning electromagnetic compatibility (EMC) must be taken for all
medical electrical equipment. The Propaq CS monitor complies with IEC EN 60601-1-
2:2001.

All medical electrical equipment must be installed and put into service in accordance
with the EMC information provided in this document and the Propaq Directions for
Use
.

Portable and mobile RF communications equipment can affect the behavior of
medical electrical equipment.

Propaq CS monitors comply with all applicable and required standards for electromagnetic
interference.

It does not normally affect nearby equipment and devices.

It is not normally affected by nearby equipment and devices.

It is safe to operate the monitor in the presence of high-frequency surgical
equipment.

However, it is good practice to avoid using the monitor in extremely close
proximity to other equipment.

Propaq CS Monitor: Guidance and Manufacturer’s Declaration—Electromagnetic Emissions

The Propaq CS monitor is intended for use in the electromagnetic environment specified below. The customer or the user
of the monitor should assure that it is used in such an environment.

Emissions Test

Compliance

Electromagnetic Environment—Guidance

RF emissions
CISPR 11

Group 1

The monitor uses RF energy only for its internal
function.Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.

RF emissions
CISPR 11

Class B

The monitor is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.

Harmonic emissions
IEC 61000-3-2

Class A

Voltage fluctuations/flicker
emissions
IEC 61000-3-3

Complies

Advertising