Creating a new calibration using two points – Vernier EasyData Guidebook User Manual

Page 47

Advertising
background image

46

EasyData Guidebook

Creating a New Calibration using Two Points

1.

From the EasyData main screen, select Setup, and
then select CH1: (or the appropriate channel).





2. Select

from the Sensor Setup screen.






3. Select

. A menu is displayed showing the different

calibration methods.

4. Select

Two Point Live….



5.

Place your sensor into the necessary condition for the
first calibration point (e.g. a force sensor with no mass
hanging from it or a pH sensor in a pH buffer). Monitor
the displayed Potential (V). When the value is stable,
select

.



6.

Enter the known value for the first calibration condition
(e.g. “0” for a force of zero N), and select

.





7.

Place your sensor into the necessary condition for the
second calibration point (e.g. a force sensor with a 200
g mass hanging from it). Monitor the displayed Potential
(V). When the value is stable, select

.