To change the type of triggering, select

CHANGE TRIGGERING

. In the

example shown, there are two types of triggering:

LabPro will trigger the start of data collection based on a change in
the data being collected. (This is called threshold triggering.)

For

MANUAL TRIGGER

, the LabPro will start collecting data when

the

START/STOP

button on LabPro is pressed.

If you choose

NONE

, no special triggering will be set.



4.

Press the number beside the type of triggering you want. If you select

NONE

,

you will return to the

ADVANCED TIME SETTINGS

screen.


If you select

MANUAL TRIGGER

, the triggering option is changed and you

will return to the

ADVANCED TIME SETTINGS

screen.


If you choose to trigger on a channel, DataMate asks you to choose the
trigger type.

INCREASING

means that data collection will not start until the

sensor readings (such as light intensity or temperature) increase
through the selected threshold (see Step 6).

DECREASING

means that data collection will not start until sensor

readings decrease through the selected threshold (see Step 6).









5.

Press the number beside the trigger type you want.


6.

Key in the number (the threshold) at which you want data collection to
begin and press

ENTER

. (Put in a threshold value in the units of the

sensor you are using such as

°C for temperature or Newtons for force.)

When the values of the data being collected reach this value, LabPro
will begin storing data.


7.

Key in the number (percent) of data you want the LabPro to prestore
and press

ENTER

. Prestore is the amount of data collected before the

threshold was reached that you want to keep (10 percent, 20 percent,
and so on). From the time the experiment begins until the threshold is
reached, LabPro collects data in a buffer. When the threshold is
reached, LabPro begins storing the data it has collected and discards the
data it collected before the threshold was reached unless a prestore
value is entered. The

ADVANCED TIME SETTINGS

screen appears.

When Manual Trigger is selected, a set of instructions will be given after
you select

START

. After selecting

START

, you can leave DataMate. Then

you can disconnect the calculator from the LabPro unit and use LabPro for
remote data collection. Data collection will begin when the Start/Stop
button on LabPro is pressed. To download data to the calculator after
remote data collection, reconnect the LabPro to the calculator and restart
DataMate. Follow the instructions on screen.