Screen explanation – Vernier DataMate Guidebook User Manual

Page 50

Advertising
background image

48

DataMate Guidebook

Screen Explanation

This screen appears when the user tries to run the Ranger
program using the TI-89/89 Titanium/92/92 Plus after using the
DataMate program.

This is caused by a conflict in some information left over in List
5. The information in the list cannot be used properly, so the
calculator gives a Dimension error. To correct this, go into the
calculator’s memory mangement and delete List 5 (L5).

This screen can appear on the TI-83 Plus. It can be caused by
running the DataMate app while the Interactive Graphing app is
loaded and turned on.

Turn off the Interactive Graphing app before running DataMate.
Also, go into memory management and check the programs
listing. There will be a program listed that has a strange character
as its name. Reset the RAM on the calculator before doing
anything else with the calculator.

When using the TI-
82 with a motion
detector and two
other analog sensors,
the data from the
sensor in Channel 2
is not collected.

The TI-82 has the capacity for only six lists, so there are not
enough lists available to collect data on all channels. When using
a motion detector, you can only use one analog sensor in Channel
1.

The sensors and
Time Graph Mode
were set up in
DataMate. Then
Triggering was set
up. When data
collection started, the
live graph did not
display.

When triggering is selected, the LabPro does not allow a live
graph.

Advertising