Technical Data

All rights reserved to make technical changes.
*

Display unit as off 100m : 1 cm

** At long range ± 5 ppm (± 0,5mm/ 100 m) plus short

range error.

Message Codes

Error message

appears beside the message number on the
display.

In case of this message switch on/off instrument
several times and check if message is still
displayed. Then call service and specify the
message displayed.

Care

Look after the optical surfaces with the same care that you
would apply to spectacles, cameras and field glasses.

Measuring accuracy

(2x standard deviation)

typ.: ± 3mm / max.: ± 5mm

Smallest unit displayed

1mm

Range

0.2m up to 200 m * **

Time for a measurement dist / trc

0.5...ca.4s / 0.16...ca.1s

∅ Laser spot (at distance)

6 / 30 / 60 mm (10 / 50 / 100 m)

Integrated telescopic viewfinder



Illumination



Multiline display



Multifunctional bumper



Time delay release



Pocket calculator



Tracking



Constant

10 values

min./ max. Tracking.



Pythagoras



Mermory (Stack)

15 Werte

Battery, Type AA, 2x 1.5V

up to 10’000 measurements

(only with Alkaline batteries!)

Splash and dust

proof

IP54 acc. IEC529: splash proof,

dust proof

Dimension and weight

172 x 73 x 45 mm, 335g

Measuring accuracy of bubble

Temperature range

Storage

Operating

-25°C to +70°C

(-13°F to +158°F)

-10°C to +50°C

(-14°F to +122°F)

Message

Code

Cause

Remedy

203

Wrong sequence with

Pythagoras meas.

Carry out measurement in

correct sequence

204

Calculation error

Repeat procedure

252

Temperature too high,

above 50°C (measuring)

Cool down instrument

253

Temperature too low,

below -10°C (measuring)

Warm up instrument

255

Receiver signal too weak,

measurement time too

long, distance <200 mm

Use target plate

measurement time >10

sec.

256

Received signal too

powerful

Use target plate (correct

side)

257

Wrong measurement,

ambient brightness too

high

Use target plate

260

Laser beam interrupted

Repeat measurement

All other messages

Call service "System"

Message

Code

Cause

Remedy