3technical data, 3 .1 technical data, 3 .2 materials used – BUCHI Recirculating Chiller F-125 User Manual

Page 12

Advertising
background image

3 Technical data

12

F-125 Manual, Version B

3

Technical data

This section introduces the reader to the Recirculating Chiller and its specifications. It contains the
technical data, requirements and performance data.

3 .1

Technical data

Technical data
Power consumption (max.)

2850 W

Supply voltage

230 VAC ± 10%

Fuse

16 AT

Output socket (max.)

2 A

Frequency

50/60 Hz

Environmental conditions
- Temperature
- Altitude
- Humidity

For indoor use only
5-40 °C
up to 2000 m
Maximum relative humidity 80% for temperatures up to 31 °C decreasing
linearly to 50% relative humidity at 40 °C

Display

digital, resolution 0.1

°C

Overvoltage category

II

Degree of protection

IP20

Pollution degree

2

Instrument dimensions
W x H x D

990 x 580 x 650 mm

Weight

100 kg

Cooling power at 15 °C

2500 W

Refrigerant

R134

Temperature range

-10 °C ... +25 °C

Hysteresis

±1

°C

Tank volume

9 l

Hose connection

13.5 mm

Pump flowrate

14 l/min

Pump pressure

2.5 bar

3 .2

Materials used

Materials used

Component

Material designation

Material code

Hazardous substances

Stainless steel, powder-coated with
polyester-epoxide

Housing

1.4301

Polyester

Lamina

PES

Glass-fiber-reinforced epoxy resin

Circuit board

Polyvinyl chloride

Cable

PVC

R134

Refrigerant

1,1,1,2-tetrafluoro-
ethane

Check MSDS

Advertising