5 Putting into operation
15
F-125 Manual, Version B
!
WARNING
Fire hazard, damage to the instrument through overheating due to inadequate air circulation.
Do not cover the instrument
•
Minimum distance from other objects must be at least 40 cm
•
!
CAUTION
Risk of minor or moderate injury due to the heavy weight of the instrument.
Utilize a second person to transport the instrument
•
Do not drop the instrument or its transport box
•
Place the instrument on a stable, even and vibration-free surface
•
Keep limbs out of crushing zone
•
5 .2
Cooling water routing
Independent of the number of coolers, the cooling water inlet has to be connected in ascending order
starting with the cooler closest to the receiving vessel. The cooling water outlet is always the upper
connector.
Example (simplified schematic)
2
3
1
1 Receiving vessel
2 Inlet (from chiller)
3 Outlet (to chiller)
The drawing shows the cooling water routing for the setup of a single cooler. Other arrangements with
single or double cooler setup can be connected accordingly.