7 Maintenance and repairs

23

F-125 Manual, Version B

7 .2

General inspection and cleaning instructions

Check the housing for visible defects (switches, plugs, enclosure etc.) and clean it regularly under
safe conditions with a damp cloth. Wipe off any splashes of aggressive chemicals immediately using
a damp cloth in order to avoid any damage being caused to the coating on the housing. Ethanol as a
cleaning agent is also possible to use.

Cleaning under safe conditions

Switch off the Recirculating Chiller and unplug the power cord.

•

Let the system acclimate to ambient temperature completely!

➡
Perform cleaning actions with a damp cloth.

•

Regularly clean lamella on front and rear side of the chiller with a damp cloth to remove dust, at

•

least once a year.

!

WARNING

Fire hazard, overheating, or triggering of the over-temperature protection function as a result of a
dirty air intake.

Clean the air intake as needed, but at least 1 x per year.

•

Hoses

Check the hoses for wear at least every six months.

•

Exchange damaged hoses.

•

Cooling media tank

Check prior to use the filling level of the Recirculating Chiller

•

Exchange the cooling liquid once a year by means of the drain valve and renew.

•

If you use a ethylene glycol / water mixture as cooling media you can drain it to the sink.

➡