8 Troubleshooting

24

F-125 Manual, Version B

8

Troubleshooting

This section helps to resume operation after a problem has occurred with the instrument which does
not require special technical training. It lists possible occurrences, their probable causes and suggests
how to remedy the problem.

8 .1

Error message display

2

3

1

1 For every error the triangle lights.

2 The error number is indicated on the

display.

3 Error 03 also includes a written message

at the top “COOLING FLUID LOW”.

8 .2

Malfunctions and their remedies

The troubleshooting table below lists possible malfunctions and errors of the instrument. The operator
is enabled to correct some of those problems or errors by him/herself. For this, appropriate corrective
measures are listed in the column “Remedy”.

Malfunctions and their remedies

Error Code

Problem

Remedy

E01

Temperature fault

Switch off the unit, let it cool down, clean air inlet and restart.
Call service if the problem persists.

E02

Temperature fault

Switch off the unit, let it cool down, clean air inlet and restart.
Call service if the problem persists.

E03

No / too little cooling liquid
Pump breakdown

Fill up cooling liquid.
Call service if the problem persists.

E04

Pressure fault

Switch off the unit; let the compressor cool down.
Call service to check the system if the problem persists.

E05

Data fault

Switch off the unit and restart.
Call service if the problem persists.

E06

Temperature fault

Switch off the unit, let it cool down, clean air inlet and restart.
Call service if the problem persists.