9. Problems and solutions

Electrode fault

T

he Compex bleeps and alternatively

displays the symbol of a pair of
electrodes and an arrow pointing to the
channel where a problem has been
detected. In the example above, the
stimulator has detected an error in
channel 1.

■

Check that electrodes are connected
to this channel.

■

Check whether the electrodes are old,
worn, and/or the contact is poor: try
using new electrodes.

■

Try using the stimulation cable on a
different channel. If the cable is still
showing a fault, replace it.

Stimulation is not producing
the usual sensation

■

Check that all the settings are correct
and ensure the electrodes are
positioned properly.

■

Change the positioning of the
electrodes slightly.

The stimulation effect causes
discomfort

■

The electrodes are beginning to lose
adhesion and no longer provide good
contact on the skin.

■

The electrodes are worn and need to
be replaced.

■

Change the positioning of the
electrodes slightly.

The stimulator is not working

■

If an error screen appears while the
equipment is in use, note the error
number (in the example in fig. 15,
error 1/0/0) and contact the
consumer service stated and
approved by Compex Médical SA.