4.

Push the Safe-Lock tensioning device so that the mounting of the

grinding bowl holder sits centrally in the U-shaped cut-out of the

longer projection.

5.

Pre-tension the setting screw (b) manually, and screw it tight using

the provided torque spanner. Place the torque spanner on the set-

ting screw (b) and turn until it clicks.

6.

Then press the clamping lever downwards. Re-tighten the addi-

tional clamping system with a hex key. The additional clamping

system may be loosened due to the clamping of the Safe-Lock

using the clamping lever.

DANGER!
After several minutes of grinding and in the cooling-down

phases, check that the tensioning devices are still firmly

connected.

If the Safe‐Lock system is correctly tensioned, the clamping

lever is automatically pulled downwards by the eccentric

when horizontal.

7.2 Grinding in inert gas with gassing lid

NOTICE!
Observe imbalance offsetting! (See Ä Chapter 6.6 ‘Mass

balance’ on page 36).