Fault description

Cause

Remedy

No mains connection

Plug in mains plug.

Main switch (10) at 0 (OFF)

Switch on main switch (I).

Device fuse

Check device fuse; fuse insert on the back side

of device 2 x 10 A T (12).

Grinding stock escapes

Tensioning device (5) loose

Check and re-tension if necessary.

Seal ring is soiled or defective

Clean or replace seal ring.

Runs unevenly with strong

vibrations

Masses balanced inadequately

Set mass balance more adequately: See

Ä

Chapter 6.6 ‘Mass balance’ on page 36.