6.5 Removing the grinding elements

1.

Lift off the lid of the collecting vessel (15) and, moving outwards

with a brush, clean the grinding stock from the outer surface of

the sieve ring (11) and from the inner edge of the collecting vessel

(8). If source material falls downward, it does not directly enter

the device; however, this should immediately be vacuumed away

with a vacuum cleaner (Ä Chapter 7 ‘Cleaning’ on page 41) after

removing the sample.

2.

Remove collecting vessel (8) with the sample.

3.

Use the vacuum cleaner to vacuum away the residual source

material

4.

Remove sieve ring (11) and rotor (10).

It is important to remove the sieve ring (11) and rotor

(10) only after removing the sample, because incom‐

pletely milled sample material could be on either

part. There is a chance that this could find its way

into the milled sample and falsify the milling results.

5.

Use a vacuum cleaner once again to vacuum away any residual

source material on the labyrinth disk.

6.

Remove labyrinth disk (5). Because the labyrinth disk (5) is

clamped and centred with an O-ring, it can be somewhat difficult

to remove.

7.

Clean parts before the next milling (Ä Chapter 7 ‘Cleaning’

on page 41).