Appendix G. Calculation of Maunal Calibration Parameters

G-5

10. The formula to corect the slope of the
line or the current calibration is simple.

Displayed weight change x
Actual weigh change

During the record keeping you find that when
you fill the vessel with 50,000 lbs of material,
the actual weight change is 47,600 lbs.

47,600 lbs/50,000lbs

*current scf_cnt = corrected scf_cnt

47,600/50,000 = .952

.952 * current scf_cnt = corrected scf_ent

This procedure will change the slope of the
line. As seen in figure G-1, the zero point will
change because of this. A new zero point will
be needed. If the amount of material in he
vessel is know, a zero-cal can be done.
Refer to page 6-6 for instructions on this
procedure.

scf_cnt = corrected scf_cnt