8 cal l 0 0, 8 cal l f 0, 88 cal 10 – A&D Weighing FC-500Si User Manual

Page 34: 88 1 pcs

Advertising
background image

display the desired calibration weight.
(For example: Using 10kg calibration weight
that actually weighs 10.002kg.)



3. Press the

ENTER

key.

The calibration weight stops blinking
and “Cal 0” appears.





4. Making sure that there is nothing on, or touching

the weighing pan, press the

ENTER

key.

When zero calibration is completed, the display
will show “Cal f”.


If you don’t need span calibration, press the
CAL switch to exit from the calibration
procedure.





5. Place the calibration weight on the weighing pan

and press the ENTER key.
When span calibration is completed, the display
will show “

Calibration End

”.



If the calibration weight is not what it should be,
an error will be displayed. Check if the weight is
correct and try again.





6. Press the CAL switch and re-attach the

calibration switch cover.
(End of the calibration procedure.)



ENTER





 



 

.

8

Cal

l

0

0

88

1

pcs

C a l i b r a t i o n Z e r o

Cala0

C a l i b r a t i o n w e i g h t ( k g )

1234567890123456789012

10.000





 



 

.

8

Cal

l

f

0

88

1

pcs

C a l i b r a t i o n S p a n

Calaf

C a l i b r a t i o n w e i g h t ( k g )

1234567890123456789012

10.000

C a l i b r a t i o n Z e r o

Cala0

C a l i b r a t i o n w e i g h t ( k g )

1234567890123456789

--------

C a l i b r a t i o n S p a n

Calaf

C a l i b r a t i o n w e i g h t ( k g )

1234567890123456789

--------





 



 

.

88

Cal

10

88

1

pcs

C a l i b r a t i o n E n d

Cal

a

C a l i b r a t i o n w e i g h t ( k g )

123456789012345678901210.000

C a l i b r a t i o n S p a n

Calaf

C a l i b r a t i o n w e i g h t ( k g )

1234567890123456789

err

If the scale will be moved to another place, set the gravity
acceleration value before calibration. The value must be of the
area where the calibration is to be done.

32

Advertising
This manual is related to the following products: