Data format – A&D Weighing FC-500Si User Manual

Page 54

Advertising
background image

Data Format

‰ Store the unit weight in use, other values (according to f-01-05) and specified item

code

Command

E ,

1 2 3 C L

F

Stores into id000123 with no item code.

E ,

1 2 3 , a b c C L

F

Stores into id000123 with item code
" b

Reply

AK


‰ ID number and item code

Command ? I D

C L

F

Reply

I D , 0 0 0 1 2 3 C L

F

A B C D E F G H I

J K L C L

F

ID number

Item code (These are “space” for no item code.)

[

” in examples below shows “Space” (20H).




‰ PCS (Count ) data

Command

? Q T C L

F

Reply

Q T

,

+ 0 0 0 0 1 2 3 4

[ P C C L

F

Stable positive data

U S

,

-

0 0 0 0 5 6 7 8

[ P C C L

F

Unstable negative data

O L

,

+ 9 9 9 9 9 9 9 9

[ P C C L

F

"E" display


‰ Weight data

Command

? W T C L

F

Reply

S T

,

+ 0 0 1

.

2 3 4 6

[ k g C L

F

Stable positive data

S T

,

-

0 0 2

.

7 2 5 5

[

l

b C L

F

Stable negative data

U S

,

-

0 0 1 2

.

3 4 6

[

l

b C L

F

Unstable negative data

U S

,

+ 0 0 0 5

.

5 9 3

[ k g C L

F

Unstable positive data

O L

,

+ 9 9 9 9

.

9 9 9

[ k g C L

F

"E" display

O L

,

-

9 9 9 9

.

9 9 9

[

l

b C L

F

"-E" display



‰ Unit weight

Command

? U W C L

F

Reply

U W ,

+ 1

.

2 3 4 5 6 7

[

[ g C L

F

U

W ,

+

0

.

2 7 2 5 3 1

[

l

b C L

F


‰ Total count

Command

? A Q C L

F

Reply

A Q ,

+ 0 0 9 9 9 9 9 9

[ P C C L

F


‰ Accumulation number

Command

? A N C L

F

52

Advertising
This manual is related to the following products: