EVCO EVF818P9 User Manual
Page 58

EVCO S.p.A.
EVF818 | Installer manual ver. 1.0 | Code 144F818E104
page 58 of 76
ESt
Configuration parameters download not completed successfully alarm.
Solutions:
-
press and release a key to restore normal display
-
download the configuration parameters again.
Main consequences:
-
the alarm output will be activated.
CEr
Configuration parameters firmware contained in EVKEY not coinciding with that of the device.
Solutions:
-
cut the device power supply off
-
check that the configuration parameters firmware contain in EVKEY coincides with that of the
device
-
download the configuration parameters again.
Main consequences:
-
the alarm output will be activated.
Erd
Configuration parameters upload not completed successfully alarm.
Solutions:
-
to restore the factory settings
-
upload the configuration parameters again.
Main consequences:
-