Screen 1: Setting the temperature

Your Grant digital ultrasonic bath is fitted with a thermostat and heaters to ensure the
temperature of the cleaning fluid is maintained at the set temperature.

Use the UP and DOWN keys to accurately set the desired temperature of the cleaning
fluid between ambient+5°C and 70°C.

During normal use, the water heaters in your Grant digital ultrasonic bath will switch
on if the cleaning fluid temperature is lower than the set temperature.
[As a safety feature, your Grant digital ultrasonic bath will only turn on when the bath
is full of fluid.]

Please note that during normal operation, ultrasonic energy will heat the cleaning fluid
by up to 15°C per hour.

Your Grant digital ultrasonic bath cannot cool the cleaning fluid. If the fluid
temperature becomes too hot, you must either let it cool down, or refill the bath with
cooler water.

SAFETY NOTICE: When the temperature is set above 50°C, the case of the Grant
digital ultrasonic bath will become hot and care should be taken when touching the
case.